Effect of different tubing and syringe material of the microfluidic systems on recovery of the dioxin mixture measured by CALUX (n = 2).
Syringes and tubing | Dioxin concentration (ng BEQ per ml) | Recoverya (%) |
---|---|---|
Polypropylene + ETFE | 0.085 ± 0.005 | 26 ± 1 |
Glass + ETFE | 0.199 ± 0.006 | 61 ± 2 |
Glass + steel | 0.212 ± 0.012 | 65 ± 4 |
Glass + Teflon | 0.219 ± 0.003 | 67 ± 1 |
The initial concentration of Dioxin in DMEM+ is 0.328 ng BEQ per ml.