Skip to main content
. Author manuscript; available in PMC: 2023 Mar 31.
Published in final edited form as: J Neural Eng. 2022 Mar 31;19(2):10.1088/1741-2552/ac59a3. doi: 10.1088/1741-2552/ac59a3

Figure 10.

Figure 10.

DyNeuMo-2 benchtop and saline test results. (A) Transient response results of RC-R network with 680 nF–1.47 μF capacitor mismatch. (B) Transient response results of RC-R network with 1–2.2 μF capacitor mismatch. (C) Saline test transient response with 1:1 electrode no surface area mismatch. (D) Saline test transient response with 2:1 electrode surface area mismatch.