Table 1.

Checklist for artifact mitigation of aDBS system with concurrent sensing and stimulation.

Signal chain element Mitigation strategy and potential trade-offs (vs)
1 IPG location to reduce ECG artifact Right chest, cranial, use of active recharge stimulation vs power consumption
2 DBS electrode impedance Minimize tissue electrode impedance mismatch, symmetric dipole arrangement
3 Frontend sensing Leakage current, polarization management, use of active recharge stimulation, stimulation pulse ‘sense blanking’
4 Digital signal processing Time vs frequency resolution trade-off in classifier, dynamic range management vs resolution floor
5 Adaptive DBS algorithm Clock synchronization and ‘algorithm blanking’ (post-stim lockout period)
6 Stimulation modulation Stimulation ramps and stimulation amplitude transitions vs response time and dynamic range of adaptive stimulation algorithm