Table 1.
Checklist for artifact mitigation of aDBS system with concurrent sensing and stimulation.
Signal chain element | Mitigation strategy and potential trade-offs (vs) | |
---|---|---|
1 | IPG location to reduce ECG artifact | Right chest, cranial, use of active recharge stimulation vs power consumption |
2 | DBS electrode impedance | Minimize tissue electrode impedance mismatch, symmetric dipole arrangement |
3 | Frontend sensing | Leakage current, polarization management, use of active recharge stimulation, stimulation pulse ‘sense blanking’ |
4 | Digital signal processing | Time vs frequency resolution trade-off in classifier, dynamic range management vs resolution floor |
5 | Adaptive DBS algorithm | Clock synchronization and ‘algorithm blanking’ (post-stim lockout period) |
6 | Stimulation modulation | Stimulation ramps and stimulation amplitude transitions vs response time and dynamic range of adaptive stimulation algorithm |