Skip to main content
. Author manuscript; available in PMC: 2022 May 12.
Published in final edited form as: Int IEEE EMBS Conf Neural Eng. 2021 Jun 2;2021:959–962. doi: 10.1109/ner49283.2021.9441336

Table 1:

Results of the ramp rate and frequency switch testing. “T&E” = trial-and-error, “Binary” = binary search for the search types.

Participant No. STN Search Type Ramp Type Iterations to Find Ramp Rate Final Imin (mA) Final Imax (mA) Ramp Rate Up (mA/sec) Ramp Rate Down (mA/sec) Clinical Stimulation Amplitude (mA) Max Tolerable Intensity for frequency Switch (mA)
1 LSTN Binary Manual 4 0.8 3 0.2188 0.2150 1.8 2.8
1 RSTN T&E Manual 5 0.9 3.8 0.0625 0.1250 2.2 2.8
2 LSTN Binary Manual 9 0.9 2.8 0.2500 0.1250 2.4 2.8
2 RSTN Binary Manual 6 1.1 3.8 0.0950 0.1250 3.2 3.5
3 LSTN Binary Varying 3 1.8 3.6 0.0625 0.1250 2.2 3.0
3 RSTN Binary Varying 2 2.0 3.9 0.1250 0.1250 2.9 1.8
4 LSTN Binary Varying 1 2.0 4.0 0.1570 0.0785 2.4 3.5
4 RSTN Binary Varying 1 2.1 4.2 0.1420 0.0711 2.6 3.6