Error % Relative ChangeBiomarkers Lung MAE LAA −950 Perc 15 Lung mass
Flash > NAEOTOM Alpha (Reference > New) −21.01% (86.39 to 68.24) −53.97% (5.59 to 2.58) −52.11% (−25.28 to −12.11) −33.59% (171.01 to 113.58)
Force > NAEOTOM Alpha (Reference > New) −22.74% (88.33 to 68.24) −68.13% (8.08 to 2.58) −64.35% (−33.96 to −12.11) −24.35% (150.13 to 113.58)