Data Processing |
|
m/z range |
1,000–8,000 |
Adjusted to the peak range in each spectrum. |
Normalize Data |
On |
|
Data Bin |
Nonlinear, 0 |
|
UniDec Parameters |
|
Charge Range |
1–60 |
These ranges can be narrowed to be targeted for a known species mass range in a low-complexity spectrum. |
Mass Range |
5,000–300,000 |
Sample Mass Every (Da) |
1 |
Decrease based on resolution and desired mass accuracy. |
Peak FWHM (Th) |
0.85 |
|
Peak Shape Function |
Gaussian |
|
Beta |
0 |
Adjusted to 50 for spectra with broad peaks. |
Charge Smooth Width |
1 |
|
Point Smooth Width |
1 |
Adjusted to 100 for spectra with broad peaks. |
Mass Smooth Width |
0 |
|
Maximum # of Iterations |
100 |
|
m/z to Mass Transform |
Smart |
|
Adduct Mass (Da) |
1.007276467 |
The charge carrier (proton) in positive mode ESI. |
Manual Mode |
Off |
Manual mode turned on to assign the +29 and +36 peaks in the POP5·RPP30 sample. |
Native Charge Offset Range |
+/− 1000 |
|
Peak Selection and Plotting |
Adjusted to label peaks in the mass spectra. |