Table A1.

Biomarker details.

Biomarkers Symbol Unit Method of Measurements/Name of the Machine and Related Information Analytical Detection Limit CHMS Reference Laboratory
Precision Target
Serum 25-hydroxyvitamin D [2,54] S-25(OH)D nmol/L Chemiluminescent immunoassay technology (DiaSorin, Ltd., Stillwater, MN, USA) 10–375 <20 nmol/L = 15%
20–100 nmol/L = 10%
>100 nmol/L = 12%
Glycated hemoglobin HbA1c % Immunoturbidimetric end-point spectrophotometry (anti-HbA1c antibody, polyhapten) on the Ortho Clinical Diagnostics Vitros 5, 1FS analyzer. %A1c is a derived test calculated from the quantitative measurements of hemoglobin and hemoglobin A1c Hb: 6.0–22.0 g/dL(component)
HbA1c: 0.2–2.5 g/dL (component)
%A1c: 4–14% NGSP
5%
Total cholesterol/HDL cholesterol ratio TC/HDL ratio Derived calculation N/A N/A
High-sensitivity C-reactive protein hs-CRP mg/L Immunoturbidimetric two-point rate reflectance spectrophotometry (monoclonal antibodies) on the Ortho Clinical Diagnostics Vitros 5, 1FS analyzer 0.10–15.0 5%
Total immunoglobulin E IgE IU/mL Two-site sandwich immunoassay using direct chemiluminescent technology on the Siemens ADVIA Centaur XP analyzer 1.5–3300 10%
Ferritin Ferritin µg/L Two-site sandwich immunoassay using direct chemiluminescent technology on the Siemens ADVIA Centaur XP analyzer 0.5–1650 10%
Hemoglobin Hb g/L Hemoglobin is a non-cyanide-based method read photometrically at 555 nm on the POCH-i100 Sysmex hematology analyzer 0–25.0 1.5%