Skip to main content
. 2022 Apr 25;14(9):1792. doi: 10.3390/nu14091792

Table 1.

Assessment of accuracy of the NMR magnesium assay.

Sample # Spiked (µM) Measured a (µM) Recovered (µM) % Recovery
1 0 589.9 ± 12.8 b -- --
2 105.1 688.5 ± 12.6 98.6 93.8
3 304.9 883.0 ± 9.5 293.1 96.1
4 609.8 1156.4 ± 5.2 566.5 92.9
5 809.5 1330.0 ± 12.0 740.1 91.4
6 1009.3 1544.1 ± 9.8 954.2 94.5
7 1503.4 2003.6 ± 41.8 1413.8 94.0
8 2008.1 2479.1 ± 36.9 1889.3 94.1

# number. a mean of four replicates ± standard deviation (SD). b measured magnesium before spiking.