Skip to main content
. 2022 Feb 14;23(5):e13562. doi: 10.1002/acm2.13562

TABLE 1.

Summary of detector systems used for commissioning the virtual cone at Tri‐Cities Cancer Center (TCCC)

Commissioning process Measurement type Detector system
Beam model adjustment HD120 MLC dosimetric leaf gap measurements Farmer chamber
Output and profile verification Virtual cone output W2, EBT3 film
Virtual cone profile EBT3 film
Absolute dose calibration ADCL‐calibrated Farmer chamber
Daily output constancy DailyQA3
Output factor constancy Virtual cone output factors Exradin A11 parallel‐plate chamber
Neutron measurements Virtual cone neutron dose Ludlum 30–7 neutron meter
Photon background Inovision 451P survey meter
Patient‐specific QA Virtual cone output and profile EBT‐XD film