. 2022 Feb 14;23(5):e13562. doi: 10.1002/acm2.13562

TABLE 7.

Measured (EBT‐XD film) virtual‐cone dose widths in the coronal plane across seven patients. The dose widths were determined from orthogonal profiles (see Figure 3) that intersect the center of the dose distribution. The measured values shown below were reduced by 2% to correct for the film image and smoothing resolution

Dose profile direction Isodose width Maximum (mm) Minimum(mm) Mean (mm) Mean difference with calculation (mm)

Left‐Right (L‐R)

50% 5.21 ± 0.07 5.10 ± 0.07 5.15 ± 0.06 0.13 ± 0.06
25% 8.66 ± 0.12 8.35 ± 0.12 8.51 ± 0.09 0.24 ± 0.09

Superior‐Inferior (S‐I)

50% 5.18 ± 0.07 5.07 ± 0.07 5.12 ± 0.06 −0.06 ± 0.05
25% 8.53 ± 0.12 8.30 ± 0.12 8.43 ± 0.09 0.12 ± 0.09