Saegerman 2021.
Study characteristics | |||
Patient Sampling |
Purpose: diagnosis of SARS‐CoV‐2 infection (mild COVID‐19 disease); to develop a clinical decision support tool for diagnosis of COVID‐19 in hospitals Design: cross‐sectional cohort study, prospective data collection Recruitment: all patients directed to the triage centres of 2 university hospital EDs Sample size: n = 2152 (573 cases) Inclusion criteria: all suspected patients directed to the triage centres (no definition of 'suspected') Exclusion criteria: patient with missing data in clinical records or missing RT‐PCR result |
||
Patient characteristics and setting |
Facility cases: positive RT‐PCR for SARS‐CoV‐2 Facility controls: negative RT‐PCR for SARS‐CoV‐2 Country: Belgium Dates: 02 March 2020‐15 June 2020 Symptoms and severity: not specified, clinical suspicion at presentation Demographics: mean age: cases 58 years, controls 52 years M%/F%: cases 48.7/51.3, controls 42.2/57.8 Exposure history: not specified |
||
Index tests |
|
||
Target condition and reference standard(s) |
|
||
Flow and timing | Index tests and RS both taken at presentation | ||
Comparative | |||
Notes | Funded by the Liège University Hospital | ||
Methodological quality | |||
Item | Authors' judgement | Risk of bias | Applicability concerns |
DOMAIN 1: Patient Selection | |||
Was a consecutive or random sample of patients enrolled? | Yes | ||
Was a case‐control design avoided? | Yes | ||
Did the study avoid inappropriate exclusions? | No | ||
Did the study avoid inappropriate inclusions? | Yes | ||
Could the selection of patients have introduced bias? | High risk | ||
Are there concerns that the included patients and setting do not match the review question? | Low concern | ||
DOMAIN 2: Index Test (All tests) | |||
Were the index test results interpreted without knowledge of the results of the reference standard? | Yes | ||
If a threshold was used, was it pre‐specified? | No | ||
Could the conduct or interpretation of the index test have introduced bias? | High risk | ||
Are there concerns that the index test, its conduct, or interpretation differ from the review question? | Low concern | ||
DOMAIN 3: Reference Standard | |||
Is the reference standards likely to correctly classify the target condition? | Yes | ||
Were the reference standard results interpreted without knowledge of the results of the index tests? | Unclear | ||
Could the reference standard, its conduct, or its interpretation have introduced bias? | Low risk | ||
Are there concerns that the target condition as defined by the reference standard does not match the question? | Low concern | ||
DOMAIN 4: Flow and Timing | |||
Was there an appropriate interval between index test and reference standard? | Yes | ||
Did all patients receive the same reference standard? | Yes | ||
Were all patients included in the analysis? | Yes | ||
Could the patient flow have introduced bias? | Low risk |