Published in final edited form as: Phys Med Biol. 2022 May 2;67(10):10.1088/1361-6560/ac6725. doi: 10.1088/1361-6560/ac6725

Table 2.

Results of ρe and SPR values calculated using the kV and MV CT calibration curves for skin, muscle, adipose, and spongiosa tissue mimicking phantoms.


Relative electron density Stopping power ratio (115 MeV)

Measured kVCT (% diff) MVCT (% diff) Measured kVCT (% diff) MVCT (% diff)

Skin 1.048 ± 0.002 1.026 ± 0.004 (−2.10) 1.051 ± 0.004 (0.29) 1.049 ± 0.002 1.055 ± 0.004 (0.62) 1.052 ± 0.004 (0.29)
Muscle 1.036 ± 0.002 1.009 ± 0.004 (−2.61) 1.038 ± 0.004 (0.19) 1.036 ± 0.002 1.038 ± 0.004 (0.13) 1.037 ± 0.004 (0.11)
Adipose 0.955 ± 0.002 0.947 ± 0.004 (−0.84) 0.963 ± 0.005 (0.84) 0.953 ± 0.002 0.978 ± 0.004 (2.58) 0.970 ± 0.005 (1.70)
Spongiosa 1.044 ± 0.002 1.067 ± 0.002 (2.20) 1.042 ± 0.003 (−0.19) 1.044 ± 0.002 1.090 ± 0.002 (4.38) 1.042 ± 0.003 (−0.22)