Fig. 7.

Fig. 7

Calibration of alcohol sensors setup at different locations in the laboratory. (a) In the solvent storage cabinet, calibration equation: Voltage = (0.29 ± 0.02) × Ethanol concentration + (1.18 ± 0.01), R2 = 0.96; (b) above the sink, calibration equation: Voltage = (0.24 ± 0.01) × Ethanol concentration + (1.16 ± 0.01), R2 = 0.97; (c) in the fume hood, calibration equation: Voltage = (0.21 ± 0.01) × Ethanol concentration + (1.51 ± 0.01), R2 = 0.92; (d) above the working bench, calibration equation: Voltage = (0.59 ± 0.03) × Ethanol concentration + (1.54 ± 0.02), R2 = 0.96. In all cases, n = 3.