Table 3.

Agreement between steps and walking measured by direct observations and by ActivPAL3™ and SENS motion® used as both single- and dual-sensor systems.

Steps Walking time
Assessment method
All gait speeds
N = 35
Slow walkers∗∗
N = 20
Fast walkers∗∗∗
N = 14
All gait speeds
N = 35
Slow walkers∗∗
N = 20
Fast walkers∗∗∗
N = 14#
Direct observation 7738 3811 3865 6397# 3758# 2446#
ActivPAL, single sensor 5096 1872 3200 3863# 1778# 2039#
SENS, single sensor 5800 2311 3311 5506# 3175# 2183#
ActivPAL, dual sensor 5096 1872 3200 3795# 1725# 2040#
SENS, dual sensor 5800 2435 3311 5506# 3175# 2182#
Percentage agreement between direct observation and ActivPAL/SENS
ActivPAL (single sensor), median % (IQR) 71.9 (39.5-80.4) 45.9 (10.7-74.7) 82.3 (75.1-90.6) 73.1 (29.3-87.6) 59.6 (9.4-82.4) 83.9 (73.1-92.1)
SENS (single sensor), median % (IQR) 78.9 (40.3-87.1) 53.33 (38.8-81.4) 83.6 (78.4-91.4) 88.9 (84.6-94.2) 89.4 (81.3-92.6) 89.4 (86.3-95.7)
ActivPAL (dual sensor), median % (IQR) 71.9 (39.5-80.4) 45.9 (10.7-74.7) 82.3 (75.1-90.6) 73.1 (29.3-87.6) 59.2 (9.3-82.4) 83.0 (73.9-95.1)
SENS (dual sensor), median % (IQR) 78.9 (40.3-87.1) 53.33 (38.8-81.4) 83.6 (78.4-91.4) 88.9 (84.6-94.2) 89.4 (81.3-92.6) 89.4 (86.3-95.7)

Number of steps; #walking time in sec; ∗∗slow walkers: <0.67 m/s; ∗∗∗fast walkers: ≥0.67 m/s. ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.