Table 8.

Relative and absolute reliability of steps, walking, and sedentary time assessed with ActivPAL3™, SENS motion®, and direct observations in older patients with hip fracture (N = 12).

Comparison Mean (SD) of both assessments (sec) Mean (SD) difference (sec) p value for difference ICC2.1 (95% CI) SEM (SEM%) MDC95 (MDC%)
Steps
ActivPAL vs. observed 124.6 (83.7) -95.8 (48.8) <0.001 0.530 (-0.089-0.868) 57.4 (46) 159.1 (128)
SENS vs. observed 139.3 (83.4) -66.3 (48.4) 0.001 0.660 (-0.081-0.910) 48.2 (35) 133.6 (96)
ActivPAL vs. SENS 91.4 (82.5) -29.5 (44.4) 0.042 0.825 (0.431-9.949) 14.4 (16) 39.9 (44)
Walking time
ActivPAL vs. observed 140.3 (47.5) -126.5 (78.5) <0.001 0.062 (-0.097-0.375) 46.0 (33) 127.5 (91)
SENS vs. observed 184.5 (54.8) -38.1 (36.8) 0.004 0.664 (-0.003-0.903) 31.8 (17) 88.1 (48)
ActivPAL vs. SENS 121.3 (51.5) -88.4 (63.2) 0.001 0.223 (-0.121-0.632) 45.4 (37) 125.8 (103)
Sedentary time
ActivPAL vs. observed 196.5 (12.1) 1.1 (2.9) 0.2 0.970 (0.903-0.991) 2.1 (1) 5.8 (3)
SENS vs. observed 192.9 (11.3) -6.0 (6.6) 0.010 0.756 0.157-0.932() 5.6 (3) 15.5 (8)
ActivPAL vs. SENS 193.5 (12.1) 7.0 (6.7) 0.004 0.716 (0.030-0.933) 6.4 (3) 17.7 (9)

ICC: intraclass correlation coefficient; SEM: standard error of measurement (SD of mean [both assessments] × √1 − ICC); MDC: minimal detectable change (SEM × √2 × 1.96, SEM% = 100 × SEM/mean). SEM and MDC indicate measurement error (in sec) at group and individual levels, respectively. ActivPAL: ActivPAL3™ monitor; SENS: SENS motion®.