Skip to main content
. 2022 May 24;22:96. doi: 10.1186/s12880-022-00814-5

Table 3.

Reproducibility of ADC Values in repeated MR exams

rs-EPI rFOV-DWI ss-EPI
ROI1
 ICCs 0.916 0.850 0.810
 BA-LA − 20 to 22 − 18 to 28.1 − 33.5 to 22.2
ROI2
 ICCs 0.797 0.595 0.442a
 BA-LA − 31 to 32.2 − 26.5 to 29 − 46.2 to 51.1
ROI3
 ICCs 0.781 0.750 0.379a
 BA-LA − 40.6 to 18.1 − 19.6 to 25.3 − 41.5 to 41.3

BA-LA (in %), the 95% Bland–Altman limits of agreements; ICCs, intraclass correlation coefficients (two-way random); rs-EPI DWI, readout-segmented echo-planar imaging diffusion-weighted imaging; rFOV, reduced field-of-view; ss-EPI, single-shot echo-planar imaging; ROI, region of interest

aP > 0.05