Figure 1.

Figure 1.

Aerosols are not generated by the Tyto cell sorter. (A) Aerosol testing impactor setup on the FACSAria II with AMS disabled and the Tyto. (B) A solution containing 1.0 μm internally fluorescent beads in the FITC channel was used for aerosol testing. (C) Representative images of coverslips analyzed after aerosol testing on the FACSAria II or Tyto with a closed (during sort) or open (postsort) door. Aerosol testing was performed in triplicate for each condition. The entire coverslip from each Tyto test was scanned and no beads were detected. AMS, aerosol management system; FITC, fluorescein isothiocyanate.