Table 1.

Results of Moist Heat Autoclave Decontamination of 3M 1860 FFRs at 115°C for 60 Minutes and 121.1°C for 30 Minutes.a

Moist Steam Decontamination Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R) (n = 1)
0.3 μm 1.0 μm 5.0 μm
Control 1860 NA NA Pass 98.62 99.66 99.89
115°C for 60 min (1 cycle) 1860 (n = 3) Pass Pass Fail (3 of 3) 89.56 99.58 100
121°C for 30 min (1 cycle) 1860 (n = 3) Pass Pass Fail (3 of 3) 88.55 98.78 99.96

a NA indicates not applicable.