Table 2.

Results of Moist Heat Autoclave Decontamination of 3M 1860 FFRs at 130°C for 2 Minutes and 130°C for 4 Minutes.a

Moist Steam Decontamination Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R) (n = 1)
0.3 μm 1.0 μm 5.0 μm
Control 1860 NA NA Pass 98.62 99.66 99.89
130°C for 2 min (1 cycle) 1860 (n = 3) Pass Pass Fail (3 of 3) 87.05 98.55 99.89
130°C for 4 min (1 cycle) 1860 (n = 3) Pass Pass Fail (3 of 3) 82.32 97.38 99.87

a NA indicates not applicable.