Table 3.

Results of Moist Heat Autoclave Decontamination of 3M 1805 FFRs at 115°C for 60 Minutes.a

Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R)
0.3 μm 1.0 μm 5.0 μm
Control 1805 NA NA Pass 86.96 ± 0.18 98.97 ± 0.13 99.92 ± 0.17
115°C for 60 min (1 cycle) 1805 (n = 3) Pass Pass Pass (3 of 3) 75.58 ± 2.50 98.04 ± 0.45 99.81 ± 0.47
115°C for 60 min (2 cycles) 1805 (n = 3) Pass Pass Pass (3 of 3) 79.08 ± 5.15 97.83 ± 0.54 99.88 ± 0.24
115°C for 60 min (3 cycles) 1805 (n = 3) Pass Pass Pass (3 of 3) 87.09 ± 0.86 98.25 ± 0.64 99.16 ± 2.53

a NA indicates not applicable.