Table 4.

Results of Moist Heat Autoclave Decontamination of 3M 1805 FFRs at 121.1°C for 30 Minutes.a

Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R)
0.3 μm 1.0 μm 5.0 μm
Control 1805 NA NA Pass 86.96 ± 0.18 98.97 ± 0.13 99.92 ± 0.17
121.1°C for 30 min (1 cycle) 1805 (n = 3) Pass Pass Pass (3 of 3) 86.96 ± 2.70 98.79 ± 0.24 99.91 ± 0.17
121.1°C for 30 min (2 cycles) 1805 (n = 3) Pass Pass Pass (3 of 3) 75.21 ± 2.98 97.96 ± 0.39 99.93 ± 0.15
121.1°C for 30 min (3 cycles) 1805 (n = 3) Pass Pass Pass (3 of 3) 59.71 ± 8.51 96.04 ± 1.05 99.83 ± 0.25

a NA indicates not applicable.