Table 5.

Results of Moist Heat Autoclave Decontamination of 3M 1870/1870+ FFRs at 115°C for 60 Minutes.a

Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R)
0.3 μm 1.0 μm 5.0 μm
Control 1870/1870+ NA NA Pass 97.71 ± 0.52 99.79 ± 0.09 100.0
115°C for 60 min (1 cycle) 1870/1870+ (n = 3) Pass Pass Pass (3 of 3) 92.41 ± 2.95 99.69 ± 0.10 99.92 ± 0.22
115°C for 60 min (2 cycles) 1870/1870+ (n = 3) Pass Pass Pass (3 of 3) 88.86 ± 3.58 99.72 ± 0.07 99.91 ± 0.23
115°C for 60 min (3 cycles) 1870/1870+ (n = 3) Pass Pass Pass (3 of 3) 87.38 ± 3.98 99.55 ± 0.19 99.83 ± 0.32

a NA indicates not applicable.