Table 6.

Results of Moist Heat Autoclave Decontamination of 3M 1870/1870+ FFRs at 121.1°C for 30 Minutes.a

Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R)
0.3 μm 1.0 μm 5.0 μm
Control 1870/1870+ NA NA Pass 97.71 ± 0.52 99.79 ± 0.09 100.0
121.1°C for 30 min (1 cycle) 1870/1870+ (n = 5) Pass Pass Pass (5 of 5) 84.48 ± 1.46 99.35 ± 0.08 99.76 ± 0.40
121.1°C for 30 min (2 cycles) 1870/1870+ (n = 5) Pass Pass Pass (5 of 5) 78.76 ± 2.44 99.34 ± 0.19 99.75 ± 0.37
121.1°C for 30 min (3 cycles) 1870/1870+ (n = 5) Pass Pass Pass (5 of 5) 59.92 ± 13.69 99.20 ± 0.30 99.69 ± 0.89
121.1°C for 30 min (5 cycles) 1870/1870+ (n = 3) Pass Pass Fail (2 of 3) NT NT NT

a NA indicates not applicable; NT indicates not tested.