Skip to main content
. 2020 Sep 1;25(3):150–156. doi: 10.1177/1535676020924171

Table 7.

Results of Moist Heat Autoclave Decontamination of ASTM Level 1 Procedural Mask at 121.1°C for 30 Minutes.a

Method Mask Type SteriGage Result ATTEST Result NIOSH Fit Test Result Mean Retention of Aerosolized Particles (%R)
0.3 μm 1.0 μm 5.0 μm
Control ASTM level 1 procedural mask NA NA NA 40.86 ± 0.62 70.74 ± 0.90 98.22 ± 0.66
121.1°C for 30 min (3 cycles) ASTM level 1 procedural mask (n = 3) Pass Pass NA 42.84 ± 1.19 68.87 ± 3.51 96.99 ± 2.30

a NA indicates not applicable.