Table 2.

K-index measurements across laboratories using Binding Site instruments and assays (B), Siemens instruments and assays (S) or a Siemens instrument coupled with a Binding Site assay (mixed). Discordant results, considering a cut-off for a positive K-index of 5.8, in bold.

Sample B1 B1 B3 B4 mixed S1 S2 S3
1 197.0 252.1 108.3 205.3 479.5 170.5 197.9 185.6
2 6.9 6.4 4.6 11.8 10.5 5.8 6.0 6.7
3 1.6 2.5 2.0 2.6 4.8 2.3 1.3 1.4
4 7.0 8.6 5.0 14.1 18.4 5.1 6.3 6.4
5 19.3 24.3 16.0 15.5 57.0 14.4 18.0 14.8
6 3.4 3.1 2.4 3.2 3.6 2.2 2.7 3.4
7 68.5 51.4 42.3 36.5 103.8 36.5 42.8 42.1
8 18.2 24.6 21.1 21.3 42.3 18.8 17.5 15.4
9 19.2 17.6 11.5 10.1 22.7 12.2 15.6 11.1
10 1.8 2.9 3.2 0.4 9.8 3.2 2.5 2.3