Evaluation of five solutions of different concentrations of H2O2 (n = 3) with three different devices with correction (WC) and without correction (WOC).

Smartphone Nº Sample Real [H2O2]
(mg/L)
WOC
(mg/L)
WC
(mg/L)
Relative Error
WOC (%)
Relative Error
WC (%)
Xiaomi Redmi 6A 1 1.36 3.60 ± 0.19 1.69 ± 0.13 164.89 24.08
2 7.82 11.47 ± 0.36 8.88 ± 0.30 46.79 13.56
3 13.6 21.22 ± 0.72 16.81 ± 0.59 56.06 23.65
4 20.4 N/A 21.29 ± 0.47 N/A 4.37
Xiaomi Mi A2 1 1.36 2.35 ± 0.06 1.95 ± 0.07 73.27 43.47
2 7.82 9.30 ± 0.27 8.92 ± 0.42 19.02 14.09
3 13.6 18.00 ± 0.34 16.96 ± 0.17 32.38 24.76
4 20.4 N/A 21.32 ± 0.47 N/A 4.54
iPhone 8 1 1.36 2.53 ± 0.13 1.75 ± 0.15 86.28 29.06
2 7.82 9.91 ± 0.21 8.84 ± 0.18 26.75 13.11
3 13.6 18.98 ± 0.32 16.97 ± 0.20 39.56 24.77
4 20.4 25.54 ± 1.10 21.21 ± 0.53 25.19 3.99