Skip to main content
. 2022 May 23;12(5):1291. doi: 10.3390/diagnostics12051291

Table 5.

The cut-off value of pre-analytical and post-analytical parameters for FFPE DNA and RNA.

Parameters Area under Curve (AUC) Std. Error Cutoff Value Sensitivity 1-Specificity
DNA Integrity Number (DIN) 0.935 0.029 3.00 1 0.152
FFPE DNA concentration (ng/µL) 0.944 0.022 5.00 1 0.5
Library concentration (nmol) (DNA) 1.000 0.000 40.65 0.99 0.00
Library concentration (nanomole) (RNA) 0.814 0.042 30.80 0.718 0.234

Table 5 The pre-analytical parameters are DIN and FFPE DNA concentration. Post-analytical parameters consist of the library concentration (nanomole) of DNA and RNA.