Skip to main content
. 2022 May 16;11(10):2798. doi: 10.3390/jcm11102798

Table 1.

Scan protocols, Siemens MAGNETOM Avanto FIT 1.5 T and Siemens MAGNETOM FreeMax 0.55 T.

Siemens MAGNETOM FreeMax 0.55 T Siemens MAGNETOM Avanto Fit 1.5 T
FLAIR (fluid attenuated inversion recovery) transversal
Field strength in T 0.55 1.5
FOV (field of view) in mm2 209 × 230 187 × 230
ST (slice thickness) in mm 3 3
SS (slice spacing) 3.6 3.6
Number of slices 40 40
PS (pixel spacing) in mm2 1.28 × 1.03 0.9 × 0.9
TR (repetition time) in msec 7780 8510
TE (echo time) in msec 96 112
TI (inversion delay) in msec 2368.8 2460
Turbo Factor 15 19
TA (time of acquisition) in min 5:28 3:26
BW (Bandwidth) 150 130
3D SWI (susceptibility weighted imaging) transversal
Field strength in T 0.55 1.5
Sequence type Multi shot 3D EPI 3D FLASH
FOV (field of view) in mm2 201 × 230 194 × 230
ST (slice thickness) in mm 3 3
Number of slices 40 48
PS (pixel spacing) in mm2 0.94 × 0.8 1.12 × 0.9
TR (repetition time) in msec 172 48
TE (echo time) in msec 100 40
Parallel imaging - GRAPPA factor 2
TA (time of acquisition) in min 2:23 2:17
BW (Bandwidth) 276 80
Single shot diffusion EPI (echo-planar imaging) transversal
Field strength in T 0.55 1.5
FOV (field of view) in mm2 220 × 220 230 × 230
ST (slice thickness) in mm 3 3
SS (slice spacing) 3.6 3.6
Number of slices 40 40
PS (pixel spacing) in mm2 1.67 × 1.67 1.44 × 1.44
b values in s/mm2 0, 1000 0, 1000
TR (repetition time) in msec 7400 6200
TE (echo time) in msec 102 103
Parallel imaging GRAPPA factor 2 GRAPPA factor 2
TA (time of acquisition) in min 4:35 2:04
BW (Bandwidth) 842 1490