. 2022 May 16;22(10):3785. doi: 10.3390/s22103785

Table 2.

The comparison of the performance of the new MWCNT-OSO3TBZ+ sensor with the performance of other potentiometric sensors for TBZ determination.

Type of the Electrode ISE with Liquid Membrane and Inner Electrolyte ISE with Liquid Membrane and Inner Electrolyte Solid-State ISE with Liquid Membrane
Ionophore (TBZ)3(PMo12O40)2 1 (TBZH2)3(PMo12O40)2 2 MWCNT-OSO3TBZ+
Slope [mV/decade of activity] 30 30 60.4 ± 0.8
Detection limit [M] 1.00 × 10−5 - 6.2 × 10−7
Useful conc. range [M] 1.0 × 10−5–1.0 × 10−2 1.0 × 10−5–1.0 × 10−2 8.6 × 10−7–1.0 × 10−3
Dynamic response [s] 40–50 (c = 10−3–10−2 M);
120–180 (c = 10−6–10−4 M)
40–50 (c = 10−3–10−2 M);
120–180 (low concentration)
8 (c = 5.0 × 10−7–1.0 × 10−4)
Working pH range 4–5 3–4 2–4
Lifetime [days] 65 60–65 90
Ref. [22] [34] This work

1 Ion pair of TBZ with 12-molybdophosphoric acid. 2 Ion pair of TBZH22+ with 12-molybdophosphoric acid.