Skip to main content
. 2021 Sep 9;69(2):250–260. doi: 10.1002/jmrs.546

Table 2.

Risk of heating of specific lead lengths at different magnetic field strengths. 8 , 30

Magnet strength RF Lead length risk
1.5T 64MHz 25‐30cm
3T 128MHz 10‐15cm
7T 298MHz 5‐7cm