. 2022 Mar 10;198(6):566–572. doi: 10.1007/s00066-022-01911-8

Table 2.

Radiation-induced ICD failures, maximum dose, and local dose rates at the specified measurement positions

Group 1 2 3 4 5
Maximum ICD dose (Gy) 3.9 1.5 3.4 1.1 0.4
Local dose rate (cGy/min) 16.8 6.6 46.8 15.6 6.0
Initial experiments (62 nonirradiated ICD)
No failuresa n = 8 (M, SJM) n = 16 (M, SJM) n = 7 (M) n = 7 (M) n = 16 (M, SJM)
Failuresa n = 2 (M, SJM) n = 3 (M) n = 3 (M)
Experiments with ICD failures at higher dose rates were repeated after 8 weeks with group 5 settings
No failuresa n = 2 (M, SJM) n = 1 (M) n = 3 (M)
Failuresa n = 2 (M)

ICD implantable cardioverter–defibrillator

a Letters in parentheses denote manufacturer: Medtronic (M); Saint Jude Medical/Abbott (SJM)