Skip to main content
. 2022 Mar 31;198(6):537–546. doi: 10.1007/s00066-022-01916-3

Table 2.

Instructions for the use of hippocampal blocking in the radiation treatment planning of hippocampus-avoidance whole brain radiation therapy with simultaneous integrated boost (HA-WBRT+SIB)

1 Define structures for optimization
HAR: the union of the two hippocampi with an isotropic margin of 7 mm
Ring_1cm: a 1 cm isotropic expansion of HAR from which HAR is subtracted
BMs: the union of all PTVBM amenable to SIB
BMs +5mm: a 5 mm isotropic expansion of BMs
PTVWB-HAR: PTVWB with HAR subtracted
PTVWB-HAR-BMs: PTVWB with HAR and BMs +5 mm subtracted
2 Set parameters for treatment fields
The isocenter is placed between the two hippocampi
Two complete arcs with a collimator angle near 0° (e.g., 5°) plus two complete arcs with a collimator angle near 90° (e.g., 85°)
In the first step the field openings are adapted to the PTVWB. In the second step the x‑jaws are closed such that the two arcs with identical collimator angle only have a small overlap between 2 cm and 3 cm (Fig. 1). This allows the MLC to block the hippocampi at all angles without also blocking larger parts of the PTV
3

Set objectives—e.g., according to the following template:

(prescription dose for PTVWB/PTVBM_1,2,3 /PTVBM_4,5,6  of 30 Gy/51 Gy/42 Gy)

Structure Type Dose [Gy] Priority gEUDa Blocking
PTVWB-HAR-BMs Upper 40.00 1000 No
Upper 30.00 1000
Lower 27.00 1000
Lower 29.20 1000
Lower 29.50 1000
Lower 28.50 1000
Upper gEUD 33.00 1000 40.0
Upper gEUD 32.00 1000 20.0
Upper gEUD 31.00 1000 10.0
GTVBM_1,2,3  Lower 51.00 1000 No
PTVBM_1,2,3  Upper 53.00 1000 No
Lower 51.00 1000
Lower 50.00 1000
Upper gEUD 52.00 1000 40.0
Lower gEUD 51.50 1000 −40.0
GTVBM_4,5,6  Lower 43.00 1000 No
PTVBM_4,5,6  Upper 43.00 1000 No
Lower 41.80 1000
Lower 41.30 1000
Upper gEUD 43.00 1000 40.0
Lower gEUD 42.50 1000 −40.0
Hippocampus left, right Upper 8.00 800 Yes
Upper 6.00 800
Mean 8.00 700
Brainstem Upper 33.00 1000 No
Lower 28.50 1000
Upper gEUD 30.30 1000 40.0
Upper gEUD 30.00 600 40.0
Chiasm Upper 31.00 1000 No
Lower 28.01 100
Upper gEUD 30.50 1000 40.0
Optic nerves left, right Upper 31.00 1000 No
Eye left, right Mean 7.00 600 No
Upper gEUD 11.00 600 10.0
Lens left, right Upper 5.00 800 No
Upper gEUD 5.00 1000 40.0
Upper gEUD 4.00 800 5.0
Ring_1 cm Upper 33.00 700 No
Lower 27.50 700
Lower 28.50 700
Lower 29.00 900
Lower 28.50 900
Upper gEUD 33.00 600 40.0
z_low (see below) Lower 29.00 1000 No
Lower 28.50 1000
Lower 28.00 1000
z_high (see below) Upper 33.00 1000 No
Upper gEUD 32.00 1000 40.0
z_out (see below) Upper gEUD 26.00 800 10.0 No
Upper gEUD 29.00 1000 40.0
4 Fine tuning
Convert 28.5 Gy isodose to structure and subtract the result from PTVWB-HAR to define the areas in which the PTV is underdosed. The corresponding volume is found as z_low in the objective template
Convert 28.5 Gy isodose to structure and crop the result from PTVWB to define the areas outside the PTV which receive a high dose. The corresponding volume is found as z‑out in the objective template
Convert 32 Gy isodose to structure and crop the result from BMs +5 mm with an additional margin of 5 mm to define the areas in which the PTV is overdosed. The corresponding volume is found as z‑high in the objective template
Optimize and repeat these steps until the result is satisfactory. Similar volumes can of course be used to optimize the dose inside the SIB volumes. However, according to our experience this is hardly necessary

WB whole brain, SIB simultaneous integrated boost, HAR hippocampus avoidance region, BMs brain metastases, RES resection cavity, PTV planning target volume, GTV gross tumor volume, MLC multileaf collimator, gEUD generalized equivalent uniform dose, gEUDa weighting parameter a of the generalized equivalent uniform dose