Table 3.
The maximum displacement and stress values of two implant models for the treatment of an unstable intertrochanteric fracture.
| Implant models | Maximum stress (MPa) |
Maximum displacement (mm) |
|||
|---|---|---|---|---|---|
| Medial cortex | Lateral cortex | Implant | Fixation models | Relative fracture surface | |
| DHS | 28.73 | 97.73 | 178.74 | 2.24 | 0.75 |
| TSFP | 68.97 | 19.58 | 105.31 | 1.57 | 0.10 |