Skip to main content
. 2022 May 25;9:911141. doi: 10.3389/fsurg.2022.911141

Table 3.

The maximum displacement and stress values of two implant models for the treatment of an unstable intertrochanteric fracture.

Implant models Maximum stress (MPa)
Maximum displacement (mm)
Medial cortex Lateral cortex Implant Fixation models Relative fracture surface
DHS 28.73 97.73 178.74 2.24 0.75
TSFP 68.97 19.58 105.31 1.57 0.10