Table 2.
Cardiac output analysis
| Stable | All | |
|---|---|---|
| Number of measurement pairs | 95 | 122 |
| CSCO: mean ± SD [l/min] | 4.21 ± 0.78 | 4.19 ± 0.82 |
| Range [l/min] | 2.95–6.65 | 2.61–6.69 |
| TDCO: mean ± SD [l/min] | 3.90 ± 0.67 | 3.83 ± 0.71 |
| Range [l/min] | 2.63–5.70 | 2.25–5.80 |
| Bias: mean ± SD [l/min] | 0.32 ± 0.51 | 0.36 ± 0.53 |
| 95% confidence interval [l/min] | 0.22–0.42 | 0.27–0.46 |
| LOA [l/min] | − 0.67–1.31 | − 0.68–1.40 |
| PE [%] | 25.15 | 26.43 |
CSCO cardiac output measured with the ClearSight system, TDCO cardiac output measured with the bolus thermodilution method, SD standard deviation, LOA limits of agreement, PE percentage error