. 2022 May 31;14(11):2744. doi: 10.3390/cancers14112744

Table 3.

Comparison of EGFR mutation detection rate in liquid biopsy between BALF and plasma.

EGFR Genotype Tissue BALF (n = 224) Tissue Plasma (n = 110)
Mutant Wild Type Mutant Wild Type
Mutant type 93 91 2 66 32 34
Wild type 131 3 128 44 6 38
Sensitivity 97.8% (91/93) (95% CI, 92.4–99.7) 48.5% (32/66) (95% CI, 35.9–61.1)
Specificity 97.7% (128/131) (95% CI, 93.5–99.5) 86.3% (38/44) (95% CI, 72.6–94.8)
PPV 96.8% (91/94) (95% CI, 90.8–98.9) 84.2% (32/38) (95% CI, 70.8–92.1)
NPV 98.4% (128/130) (95% CI, 93.5–99.5) 52.7% (38/72) (95% CI, 46.2–72.6)
Concordance rate 97.7% ((91 + 128)/224) (95% CI, 94.8–99.2) 63.6% ((32 + 38)/110) (95% CI, 53.9–72.6)

Abbreviations: BALF, bronchoalveolar lavage fluid; CI, confidence interval; EGFR, epidermal growth factor receptor; PPV, positive predictive value; NPV, negative predictive value.