Table 1.
Examples of nanomaterial-based NT detection systems.
Detection System | Detected NT | Performance Indicators | Ref. |
---|---|---|---|
Glutamate oxidase entrapped in a chitosan matrix cast onto the microelectrode surface (i.e., platinum wire covered with poly-o-phenylenediamine) and coated with ascorbate oxidase | Glutamate | Sensitivity: 0.097 ± 0.001 nA μM−1 Linearity range: 5–150 μM Detection limit: 0.044 μM |
[146] |
Monolithic platform based on carbon-nanotube field-effect transistors | Glutamate | Linearity range: 250–500 μM Detection limit: 3 μM |
[147] |
Perovskite nickelate-Nafion (i.e., polymeric ion-permeable membrane) heterostructure | Glutamate | Sensitivity: 0.327 ± 0.07 nA μM−1 mm−2 Linearity range: 1–700 μM Detection limit: 16 nM |
[148] |
Reduced graphene oxide-based field-effect transistor biosensor functionalized with mGluR | Glutamate | Linearity range: 1 fM–100 pM Detection limit: 1 fM |
[149] |
Enzyme-free electrochemical sensor based on graphene oxide modified gold electrode | GABA | Linearity range: 250 nM–100 μM Detection limit: 98 nM |
[150] |
nanoITIES (interface between two immiscible electrolyte solutions) pipet electrodes | GABA | Linearity range: 0.25–1.0 mM Detection limit: 22.4 μM |
[151] |
Gold nanoparticles-zinc oxide nanocone arrays/graphene foam electrode | Dopamine | Sensitivity: 4.36 μA mM−1 Detection limit: 0.04 μM |
[152] |
Label-free luminescent NaGdF4:Tb nanoparticles | Dopamine | Linearity range: 0–10 μM Detection limit: ~30 nM |
[153] |
Supramolecular β-cyclodextrin functionalized gold nanoclusters | Dopamine | Linearity range: 100.0 nM–80.0 μM Detection limit: 20.0 nM |
[154] |
Graphite screen-printed electrodes modified by a nanocomposite made of polyaniline and gold nanoparticles | Dopamine | Linearity range: 1–100 μM Detection limit: 0.86 μM |
[155] |
Nanocomposite platform based on graphene oxide/chitosan modified screen-printed electrode | Serotonin | Sensitivity: 0.05 μA mM−1 Detection range: 10 nM–100 μM Detection limit: 3.2 nM |
[156] |
Gold-nanorattles-reduced graphene oxide nanocomposite coated onto the gold nanoparticles deposited glassy carbon electrode (GCE) | Serotonin | Linear dynamic range: 3 × 10−6–1 × 10−3 M Detection limit: 3.87 (±0.02) × 10−7 M |
[157] |
Graphite-paste electrode modified with nanoparticles (i.e., Fe3O4@Au@SiO2) coated with molecularly imprinted polymer | Serotonin | Linearity range: 0.01–1000 μM Detection limit: 0.002 μM |
[158] |
GCE coated with a biofilm of graphite, nanodiamonds, and gold nanoparticles anchored in casein | Serotonin | Sensitivity: 0.18 μA mM−1 Linear dynamic range: 0.3–3.0 μM Detection limit: 0.1 μM |
[159] |
Platinum nanoparticles coated with molecularly imprinted silica drop-cast onto a GCE | Serotonin | Linearity range: 0.05–80 μM Detection limit: 0.02 μM |
[160] |
Laccase modified GCE coated with graphene quantum dots | Epinephrine | Sensitivity: 2.9 μA mM−1 cm−2 Linearity range: 1–120 × 10−6 M Detection limit: 83 nM |
[161] |
Tetrahexahedral gold-palladium core-shell nanocrystals on reduced graphene oxide nanosheets | Epinephrine | Linear detection range: 0.001–1000 μM Detection limit: 0.0012 μM |
[162] |
GCE modified with chemically reduced graphene oxide nanosheets | Epinephrine | Two different linearity ranges: 10–300 and 400–1300 μM Detection limit 1.6 μM |
[163] |
Graphite screen-printed electrode modified with a nanocomposite of magnetic Fe3O4@SiO2 nanoparticles and carbon nanotubes | Norepinephrine | Linearity range: 0.5–400 μM Detection limit: 0.2 μM |
[164] |
GCE modified with carbon nanotubes and magnetic nanoparticles of cobalt ferrite | Norepinephrine | Linearity range: 0.16–1.91 mM Detection limit: 0.76 μM |
[165] |
Copper-palladium core-shell nanostructures on pencil graphite substrate | Histamine | Sensitivity: 0.082 μA μM−1 cm−2 Detection limit: 3.2 ± 0.1 nM |
[166] |
Nickel-based metal-organic framework crystals and multi-walled carbon nanotubes modified GCE | Histamine | Sensitivity: 0.19 μA μM−1 Linearity range: 1.00–160.00 μM Detection limit: 0.41 μM |
[167] |
Fiber-optic surface plasmon resonance (SPR)-based biosensor covered with multilayers of silver metal and tantalum (V) oxide nanoflakes functionalized with acetylcholinesterase enzyme | Acetylcholine | Sensitivity: 8.709 nm/μM Detection limit: 38 nM |
[168] |
Acetylcholinesterase and choline esterase co-immobilized on platinum nanoparticles and metallic organic framework modified gold electrode | Acetylcholine | Linearity range: 0.01–500 μM Detection limit: 0.01 μM |
[169] |
Acetylcholinesterase and choline esterase co-immobilized over a gold electrode coated with a nanocomposite layer of multi-walled carbon nanotubes and reduced graphene oxide | Acetylcholine | Linearity range: 0.1–100 μM Detection limit: 0.1 μM |
[170] |
Enzyme-free electrochemical sensor based on spinel-type copper cobaltite nanoplates | Acetylcholine | Linear dynamic range: 0.2–3500 μM Detection limit: 30 nM |
[171] |