Table 1.

This table shows the applied imaging sequences according to ASTM F2119

Sequences
Spin echo Gradient echo
TE [ms] 20 15
TR [ms] 500 500
Flip angle [°] 90 30
Matrix size 256 × 256 256 × 256
Slice thickness [mm] 3 3
Pixel bandwidth [Hz/px] 130 130
FOV [mm2] 200 × 200 200 × 200

In accordance with ASTM F2119, no geometric distortion correction or accelerated imaging acquisitions methods were used