Table 5.

Comparison of NECR and scatter fraction (SF) of different preclinical PET scanners. Data are taken from [515]

System Mouse phantom Rat phantom
NECR peak (kcps) Activity (MBq)b SF (%) NECR peak (kcps) Activity (MBq) SF (%)
SAFIRa 799 537 10.9 121 624 17.8
Hyperion IID 407 46 13
NanoScan 406 30 17.3 119 25 34
MuPET 1100 57 11.9 352 65 28.0
Inveon 1670 131 7.8 592 110 17.2
IRIS 185 14 15.6 40 10 22.4
ClearPET 73 18 31
Mosaic HP 555 92 5.4 244 87 12.7
LabPET 8TM 279 82 15.6 94 91 29.5
microPET R4 168 91 18 89 81 28
Xtrim-PET 113.2 17 12.5 82.8 15 25.8
Bruker 486 23 240 23

aFor SAFIR, the highest NECR values at the highest measured activities are reported as the NECR peaks are not reached

bActivity at which the NECR peak is measured