Table 2. Repeatability and within-lab (total) imprecision for the hs-TnI assay on the Atellica IM analyser at different sites.

Site Sample type Mean Conc. (ng/L) Repeatability Within-Lab (total)
SD CV(%) SD CV(%)
A SP1
SP2
LT
L1
L2
11.9
26.4
99.2
260.9
5863.9
0.5
0.7
1.6
4.3
66.9
4.1
2.6
1.6
1.7
1.1
0.9
1.5
4.4
11.9
186.9
7.6
5.5
4.4
4.6
3.2
B SP1
SP2
LT
L1
L2
10.7
25.0
95.0
252.8
5717.0
0.4
0.7
1.5
4.4
67.0
3.6
2.9
1.6
1.7
1.2
0.7
0.9
3.8
10.9
154.9
6.8
3.4
4.0
4.3
2.7
C SP1
SP2
LT
L1
L2
11.0
25.2
93.0
245.6
5577.7
0.4
0.6
1.8
5.5
78.8
3.6
2.2
2.0
2.2
1.4
0.6
0.8
3.0
7.6
118.7
5.5
3.0
3.3
3.1
2.1
D SP1
SP2
L1
L2
L3
11.2
25.4
40.3
3684.1
19,228.1
0.5
1.0
0.7
57.4
447.5
4.7
3.8
1.7
1.6
2.3
0.8
1.0
1.0
67.6
675.3
6.7
3.9
2.5
1.8
3.5
hs-TnI – high sensitive troponin I. SD – standard deviation. CV – coefficient of variation. Conc. – concentration.