Table 3. Correlation and Passing-Bablok regression results for hs-TnI assays on the Atellica IM 1600 Analyser and established systems.

Site Comparative hs-cTn assay N Measuring range
(ng/L)
(Atellica IM)
Slope 95% CI of slope Intercept
(ng/L)
95% CI of intercept Correlation coefficient
A Advia Centaur XP 39 11.84–18,397 0.88 0.86–0.90 0.77 -0.33–2.61 0.997
B Advia Centaur XP* 87 2.21–15,989 1.02 1.00–1.03 0.98 0.67–1.41 0.999
C Advia Centaur XP§ 117 2.19–7,223 1.06 1.04–1.08 0.87 0.66–1.10 0.999
D Advia Centaur XP§ 72 1.53–20,212 1.05 1.03–1.06 -2.31 -2.70–(-1.63) 0.999
B Architect STAT* 91 1.34–15,989 0.99 0.88–1.14 1.46 0.28–2.71 0.964
C Dimension Vista 121 1.49–7,223 1.04 1.00–1.07 -1.37 -1.63–(-0.76) 0.984
A Dimension EXL* 40 11.84–19,566 0.97 0.93–1.01 -0.24 -6.12–9.48 0.998
CI – coefficient of variation. hs-Tn – high sensitive troponin. *Intercept 95% CI includes value zero and slope 95% CI includes value one; Intercept 95% CI does not include value zero and slope 95% CI includes value one; Intercept 95% CI includes value zero and slope 95% CI does not include value one; §Intercept 95% CI does not include value zero and slope 95% CI does not include value one