Table 5. Results of the Atellica IM 1600 turnaround time (TAT) for the hs-TnI assay during testing of 1461 routine and 13 hs-TnI STAT samples.

Total TAT TAT
from registration
to entrance in IOM*
TAT
from IOM to centrifuge (connected)
TAT
from entrance in IOM to sample aspiration in Atellica IM analyser
TAT
in Atellica IM analyser (inside the analyser): from bar code reading to result
Mean (min.) 44.4 6.8 19.3 27.2 10.7
SD (min.) 12.4 5.5 3.0 8.0 0.4
P90 (min.) 63.1 14.0 23.6 38.4 11.2
CV (%) 28 80 16 29 4
Minimum 18.2 0.0 13.5 7.0 10.1
Maximum 74.7 29.5 29.8 44.8 15.8
*IOM – input/output module from the APTIO Automation Solution (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) used with the Atellica IM analyser. TAT – Turnaround time. SD – standard deviation. CV – coefficient of variation. Min – minutes. STAT – short turnaround time. hs-TnI – high sensitive troponin I. P90 – 90th percentile.