Table 5.

Representative selection of the patient cohort, each with a different valve explanted

Valve type Valve size Perceval implanted ID of explant valve ID of implant valve
St Jude Epic 21 L 19 25
St Jude Trifecta 23 XL 21 27
Carpentier-Edwards 19 M 18 23
Carpentier-Edwards Magna Ease 23 L 22 25
Valve type Valve size Perceval implanted ViV TAVR size for explant (S3) ViV TAVR size for implant (S3) ViV TAVR size for explant (Evolut) ViV TAVR size for implant (Evolut)
St Jude Epic 21 L 20 23/26 23 26/29
St Jude Trifecta 23 XL 23 26/29 26 29
Carpentier-Edwards 19 M 20 23 23 26
Carpentier-Edwards Magna Ease 23 L 23 23/26 26 26/29

The ID of the Perceval implanted is much larger than that of the explanted valve.19 Sutureless valve implantation consistently allows for a larger Sapien 3 or Evolut ViV TAVR option compared with the explanted valve.19ID, Internal diameter; L, large; XL, extra-large; M, medium; ViV TAVR, valve-in-valve transcatheter aortic valve replacement; S, small.