Skip to main content
. 2022 May 24;60(6):e00098-22. doi: 10.1128/jcm.00098-22

TABLE 1.

Overall performance of the Reveal AST system

Parameter or detail Performance of Reveal AST against:
Sensititre Vitek 2
Parameter, % (no. positive/total no.)
 EAc 98.0 (2,129/2,173) 97.0 (1,482/1,528)
 CAc 96.3 (2,174/2,258) 96.2 (1,554/1,615)
 mE 3.5 (78/2,258) 3.3 (54/1,615)
 ME 0.3 (5/1,889) 0.3 (4/1,342)
 VME 1.3 (4/313) 1.3 (3/232)
Study set details
 No. of species 7 7
 No. of antimicrobials 24 17
 No. of strainsa 101 104
 No. of total strain-drug pairsb 2,258 1,615
 No. of S strain-drug pairs 1,889 1,342
 No. of I strain-drug pairs 56 41
 No. of R strain-drug pairs 313 232
 % R + I 16.3 16.9
a

Sensititre results were not available for 3 strains.

b

The total number of strain-drug pairs is the number that was evaluated against Sensititre/Vitek 2 and excludes pairs that show intrinsic resistance (see Table 2 for details).

c

ESBL screen results are included in the CA calculations but not the EA.