. Author manuscript; available in PMC: 2022 Jun 15.
Published in final edited form as: Med Phys. 2022 Feb 25;49(5):3159–3170. doi: 10.1002/mp.15542

Table 1:

Imaging parameters of the enrolled patients.

Original 4D-MRI T1w 3D MRI T2w 3D MRI
No. of Patients 39 27 27
Contrast-enhanced No Yes No
Acquisition mode Free breathing Free breathing Gated
Sequence TWIST-VIBE Star-VIBE TSE
Fat suppression N/A SPAIR SPAIR
Turbo factor N/A N/A 43
Flip angle (°) 5 9 72
Echo trains per slice N/A N/A 6
Parallel imaging (factor) CAIPIRINHA (4) N/A GRAPPA
Acceleration factor (PE) 2 N/A 3
Partial Fourier 6/8 7/8 N/A
TR (ms) 3.44 2.83 1090.0
TE (ms) 1.23/2.45 1.48 84.0
Bandwidth (Hz/pixel) 1420 820 781
Matrix size 160 × 128 × 64 320 × 320 × 72 256 × 256 × 40
Voxel size (mm) 2.7 × 2.7 × 2.7 1.2 × 1.2 × 3.0 1.5 × 1.5 × 5.0
Acquisition time 0.69 s per 3D frame 173 s 60 s
No. of frames 72 N/A N/A

3D: Three-dimensional; 4D: Four-dimensional; CAIPIRINHA: Controlled aliasing in parallel imaging results in higher acceleration; GRAPPA: Generalized autocalibrating partial parallel acquisition; MRI: Magnetic resonance imaging; SPAIR: Spectral attenuated inversion recovery; T1w: T1-weighted; T2w: T2-weighted; TE: Echo time; TR: Repetition time; TSE: Turbo spin echo; TWIST: time-resolved imaging with interleaved stochastic trajectories; VIBE: volumetric interpolated breath-hold examination.