Skip to main content
. 2022 May;10(10):588. doi: 10.21037/atm-22-1642

Table 5. Assessment results for the flagging ability of abnormal cells in BC-7500 CRP.

Flagging TP FP TN FN Sensitivity (95% CI) Specificity (95% CI) Efficiency (95% CI)
Blast 31 123 639 2 93.9% (80.4–98.3%) 83.9% (81.1–86.3%) 84.3% (81.6–86.6%)
Immature Gran 284 181 304 26 91.6% (88.0–94.2%) 62.7% (58.3–66.9%) 74.0% (70.8–76.9%)
NRBC 104 17 658 16 86.7% (79.4–91.6%) 97.5% (96.0–98.4%) 95.8% (94.2–97.0%)
WBC total 362 184 233 16 95.8% (93.2–97.4%) 55.9% (51.1–60.6%) 74.8% (71.7–77.7%)

CRP, C-reactive protein; NRBC%, nucleated red cells percentage; WBC, white blood cell; 95% CI, 95% confidential interval; TP, true positive; FP, false positive; TN, true negative; FN, false negative.