Table 7.

Summary of standard mp-MRI protocol used in image acquisition.

Pulse sequence Dual echo T1W GRE T2W TSE/FSE Volume interpolated T1W 3D GREb Diffusion weighted imagingc
2D GRE 3D GREa Single shot TSE/FSE 3T 1.5T Single shot echo-planar imaging
3T 1.5T
Physiology Breath hold Breath hold Breath hold Respiratory triggered Breath hold Breath Hold
Breath hold
Duration 21 s 16 s 20 s 3 to 4 min 20 s 21 s
22 s
Fat suppression N/A N/A N/A N/A Chemical or spectral inversion recovery Spectral inversion recovery
TE (IP/OP)d; TR (ms) (4.6/2.3);160 to 180 (2.5/1.3); 5.5 and (2.2/1.1); 4.0 (4.6/2.3); 7.6 83 to 88; 1030 1.7 to 2.5; 4.0 to 4.5 1.4;4.3 60.8 to 74; 2075 to 4600
Flip angle (deg) 70 10 to 12 10 180 10 to 12 10 to 12 90
Bandwidth (Hz) 260 700 313 450 325 to 460 488 250 to 1446
Number of excitations 1 0.7 to 1 1 Half-Fourier 1 1 2
Acceleration factor 2 2 1 1 2 2 2
2
Matrix size 256/320 × 134/152 294 × 224 192 × 320 170 × 256 256 × 320 132 × 320 130 to 38; 96 to 75
Field of view (cm) 25 × 35 25 × 35 25 × 35 25 × 35 25 × 35 25 × 35 40 to 380; 28 to 75
Slice thickness (mm) 5 to 6 3 to 4 3 to 5 5 2.5 to 4 2.5 to 4 6
a

Imaging was performed on clinical 1.5 Tesla or 3 Tesla systems.

b

VIBE (Siemens Healthcare), THRIVE (Philips Healthcare), LAVA (General Electric Healthcare).

c

Diffusion weighted imaging performed with two b values (0 and 600  mm2/s) with ADC map automatically derived.

d

IP = in phase, OP = opposed phase.