. 2022 May 28;94(23):8156–8163. doi: 10.1021/acs.analchem.1c05206

Table 2. Sensitivity (LoD), Accuracy (% Recovery), and Precision (% CV) of TxB and GDH Assays, as Determined from Raw Bioluminescence (RLU; Figures 4 and S11C,D).

  intra-assay interassay
TxB
sensitivity (LoD) 44 fM 190 fM
quantifiable range 0.1–1000 pM 1–100 pM
% recovery 88–105% 99–101%
% CV 3–25%a 11–20%
GDH
sensitivity (LoD) 4.5 pM 14 pM
quantifiable range 0.01–10 nM 0.1–10 nM
% recovery 100% 101–102%
% CV 1–10% 19–24%a
a

%CV precision metrics > 20% only at the limit of quantification.