Skip to main content
. 2022 May 12;94(23):8136–8145. doi: 10.1021/acs.analchem.1c04450

Table 1. Precolumn Pressure and Dilution Comparisona.

precolumn ACN [%] flow. [mL/min] temp. [°C] ID [mm] length [mm] p. size [μm]
bpC1 1.0 2.20 30 4.6 5 2.7
cpC2 1.2 1.70 30 3.0 5 1.9
dpC3 1.5 1.35 30 2.1 5 1.7
epC4 2.3 0.80 60 4.6 30 2.7
fpC5 4.5 0.25 60 3.0 30 1.7
gpC6 6.1 0.11 60 2.1 100 3.5
hpC7 8.3 0.00 60 2.1 50 1.7
a

Comparison of different precolumns used for peptide trapping with the mD-UPLC-MS/MS system. The pressure was measured by the 3D-pump during the analysis of the main peak fraction of Herceptin (trastuzumab, 50 μg injection). The mD-UPLC-MS/MS system was operating in the single-enzyme digestion mode with a trypsin IMER installed. The listed flow rates (Flow.) represent the 4D-pump flow rate for dilution excluding the 2D-pump 0.05 mL/min (50% ACN) and 3D-pump 0.25 mL/min (digestion buffer) flow rates. For each column, the highest 4D-pump flow rate is listed before the 3D-pump exceeds the pressure limit of the 3D-trypsin column (<170 bar). In addition, the calculated acetonitrile concentration while peptide trapping on the 4D precolumn is listed (ACN). For the small trapping columns with a length of 5 mm, the temperature was set to 30 °C for optimal trapping performance. For longer columns, the temperature (temp.) was set to 60 °C because of the high backpressure at low temperatures. Additionally, the inner diameter (ID), length, and particle size (p. size) are listed for each column.

b

pC1: Agilent InfinityLab Poroshell 120 SB-C18 Fast Guards.

c

pC2: Agilent InfinityLab Poroshell 120 SB-C18 Fast Guards.

d

pC3: Waters ACQUITY UPLC BEH C18 Precolumn.

e

pC4: Agilent InfinityLab Poroshell 120 SB-C18 Column.

f

pC5: Waters ACQUITY UPLC Peptide BEH C18 Column.

g

pC6: Waters XSelect Peptide CSH C18 Column.

h

pC7: Waters ACQUITY UPLC Peptide BEH C18 Column.