Table 1. Precolumn Pressure and Dilution Comparisona.
| precolumn | ACN [%] | flow. [mL/min] | temp. [°C] | ID [mm] | length [mm] | p. size [μm] |
|---|---|---|---|---|---|---|
| bpC1 | 1.0 | 2.20 | 30 | 4.6 | 5 | 2.7 |
| cpC2 | 1.2 | 1.70 | 30 | 3.0 | 5 | 1.9 |
| dpC3 | 1.5 | 1.35 | 30 | 2.1 | 5 | 1.7 |
| epC4 | 2.3 | 0.80 | 60 | 4.6 | 30 | 2.7 |
| fpC5 | 4.5 | 0.25 | 60 | 3.0 | 30 | 1.7 |
| gpC6 | 6.1 | 0.11 | 60 | 2.1 | 100 | 3.5 |
| hpC7 | 8.3 | 0.00 | 60 | 2.1 | 50 | 1.7 |
Comparison of different precolumns used for peptide trapping with the mD-UPLC-MS/MS system. The pressure was measured by the 3D-pump during the analysis of the main peak fraction of Herceptin (trastuzumab, 50 μg injection). The mD-UPLC-MS/MS system was operating in the single-enzyme digestion mode with a trypsin IMER installed. The listed flow rates (Flow.) represent the 4D-pump flow rate for dilution excluding the 2D-pump 0.05 mL/min (50% ACN) and 3D-pump 0.25 mL/min (digestion buffer) flow rates. For each column, the highest 4D-pump flow rate is listed before the 3D-pump exceeds the pressure limit of the 3D-trypsin column (<170 bar). In addition, the calculated acetonitrile concentration while peptide trapping on the 4D precolumn is listed (ACN). For the small trapping columns with a length of 5 mm, the temperature was set to 30 °C for optimal trapping performance. For longer columns, the temperature (temp.) was set to 60 °C because of the high backpressure at low temperatures. Additionally, the inner diameter (ID), length, and particle size (p. size) are listed for each column.
pC1: Agilent InfinityLab Poroshell 120 SB-C18 Fast Guards.
pC2: Agilent InfinityLab Poroshell 120 SB-C18 Fast Guards.
pC3: Waters ACQUITY UPLC BEH C18 Precolumn.
pC4: Agilent InfinityLab Poroshell 120 SB-C18 Column.
pC5: Waters ACQUITY UPLC Peptide BEH C18 Column.
pC6: Waters XSelect Peptide CSH C18 Column.
pC7: Waters ACQUITY UPLC Peptide BEH C18 Column.