Skip to main content
. 2022 May 31;24(6):373–380. doi: 10.1089/dia.2022.0011

Table 4.

Sensor Accuracy by Diabetes Type and Insulin Regimen

Placement Diabetes type and insulin regimen No. of subjects No. of matched pairs %15/15 (%) %20/20 (%) %30/30 (%) MARD (%)
Arm (N = 308) Type 1 252 31,948 89.8 95.4 98.8 8.2
Type 2—IIT 46 6005 88.8 94.7 98.7 8.0
Type 2—NIIT 10 1240 87.7 96.0 99.0 9.2
Abdomen (N = 311) Type 1 253 31,109 85.0 92.5 97.9 9.3
Type 2—IIT 48 6260 87.6 95.8 99.1 8.6
Type 2—NIIT 10 1212 87.8 97.2 99.6 7.6

IIT, intensive insulin therapy; NIIT, nonintensive insulin therapy.